LIQUID CONNECTIONS
SS316 1/4" Swagelok-type compression fitting
|
OPERATION MODE
- Autonomous: all unit functions are controlled from the front panel, no external PC is needed.
- Up to seven parameters are continuously recorded to the detector non-volatile memory.
- Up to four parameters can be selected at any time by user to be plotted as real-time graphs on the color LCD.
|
LIGHT SOURCES
- 280nm and/or 250nm UV LEDs, followed by narrow bandpass diffraction filters
- Spectral bandwidth: 10nm
- Wavelength accuracy: ± 3nm
- Current through UV LEDs: ≈ 10% of the maximum allowed for the corresponding UV LED. The operating current reading is available on the detector's screen.
- Thermally compensated to minimize long-term drift
- Typical lifetime: 15 years of non-stop operation
|
COMBINED UV/CONDUCTIVITY FLOW CELL
- A single REACH Devices wide-range embedded flow cell is implemented, US Patent 8570521.
- Total flow cell volume: 400μl
- Flow rates: 1 to 1000 ml/min, no split - all eluants travel through the flow cell via the same path.
- Maximum operating pressure: 1000 psi (68 bar)
- Wetted materials: quartz, sapphire, PEEK, titanium, FFKM (Kalrez)
- Solvents compatibility: all except HF, concentrated
H2SO4,
concentrated HNO3, and molecular halogens
- Long term pH stability: 0-13
- Hydrodynamic resistance: low, see graph
|
UV MONITORING
- Range: 0.001 – 800 AU (Absorption Units)
- Resolution: 0.0001 AU
- Baseline noise: < 0.0002 AU
- Drift: 0.0003 AU/hr at constant temperature (dry flow cell)
|
CONDUCTIVITY MONITORING
- Two bright platinum electrodes
- 0.01-150 mS/cm range
|
REFRACTIVE INDEX MONITORING
- Measurement type: absolute
- Measuring wavelength: 280nm
- Range: 1.2000 to 1.7000
- Resolution: 0.0001
- Thermally compensated measurements
|
pH FLOW CELL
- Total flow cell volume: 500μl
- Flow rates: 1 to 1000 ml/min, no split - all eluants travel through the flow cell via the same path.
- Maximum operating pressure: 5 psi with the pH electrode, 15 psi with the dummy electrode
- Wetted materials: stainless steel 316, Viton, PEEK
- Solvents compatibility is limited by the pH electrode
- Hydrodynamic resistance: 3 psi back-pressure at 1 L/min for water at 25
oC
|
pH MONITORING
- Electrical connector: BNC
- Input current: <1pA
- Input voltage range -1500 to +1500 mV, will accept any type of potentiometric pH or Ion Selective electrode
- Electrode resistance measurement range: 1-5000 MOhm
|
POWER OUTAGE
The recorded data is not corrupted. The detector will automatically resume data recording after the power is back on.
|
ACQUISITION RATES (non-stop run times)
- 5Hz (2 hour 38 min)
- 2Hz (6 hours 37 min)
- 1Hz (13 hours 14 min)
- 0.5Hz (26 hours 28 min)
- 0.2Hz (66 hours 10 min)
- 0.1Hz (132 hours 20 min)
|
DATA TRANSFER
By user's command to USB drive. No data acquisition occurs during data transfer.
|
COMMUNICATION INTERFACE
- 4 independent analog outputs in 0-1V range (see manual for transfer functions), corresponding to
(a) UV 250 nm, (b) UV 280 nm, (c) refractive index, (d) conductivity channels.
The output (a) can be reassigned by a user to provide data from any channel, including pH, and used to provide
the respective analog signal for contemporary fraction collectors.
- One digital input pin for fraction count pulse (e.g. from a fraction collector).
Detector can react to intermittent mechanical contact closure and TTL level signal.
- Two digital outputs which can be programmed by REACH Devices as requested by the user
- All input and output channels are galvanically isolated from mains ground and
detector's internal circuitry ground for safety and noise resistance.
- Up to two relays which can be set by the user to react on change in UV absorption, pH or conductivity.
|
CUSTOM FIRMWARE UPLOAD
In the field from a flash drive
|
DISPLAY
- Dimensions: 5.7” diagonally
- Resolution: QVGA 320x240
- Back-light: white LED
- Protection: hermetically covered with 2 mm glass – the assembly is solvent-resistant.
|
CASE MATERIAL
Anodized titanium, covered with 1 micrometer Teflon –
complete resistance to any solvent spills.
|
DIMENSIONS
7.5” x 5.25” x 5.05” (190 x 133 x 128 mm)
|
WEIGHT
- No valves: 4 lb (1.8 kg)
- 1-8 valves: 4.2 - 5.5 lb (1.9 - 2.5 kg)
|
POWER REQUIREMENTS
- Embedded internal (USA, Asia, and most other regions) or included external (Canada, Europe) power supply
- Frequency: 50-60Hz
- Voltage: 90-240V
- Current rating: 0.1A for 90-120V, or 0.06A for 220-240V; 10W max power draw
|