RD4-pH: 250-280-pH-Cond, Industrial

Industrial Capacity UV, pH and Conductivity Monitor and Controller for liquid chromatography

GeneralSpecificationsOptionsFraction RouterUser Manual

Operation mode

Autonomous: all unit functions are controlled from the front panel, so no external PC is needed.

Data presentation

Four real-time graphs (one for each data channel) are plotted on a color LCD.

Light sources

  • 250nm and 280nm UV LEDs, followed by narrow bandpass diffraction filters
  • Spectral bandwidth: 9nm
  • Wavelength accuracy: ± 2nm
  • Current through UV LEDs: ≈ 20% of the maximum allowed for the corresponding UV LED. The operating current reading is available on the detector's screen.
  • Light sources are kept at a constant temperature by a Peltier element to minimize long-term drift
  • Typical lifetime: 15 years of non-stop operation

Combined UV/Conductivity flow cell

  • A single wide-range embedded flow cell is implemented, patent pending.
  • Total volume: 400μl
  • Flow rates: 0.1 to 1500ml/min, no split - all eluents travel through the flow cell via the same path.
  • Maximum operating pressure: 2500 psi (17 bar)
  • Wetted materials: sapphire, titanium
  • Solvents compatibility: any (except HF; see manual)
  • pH stability (long term): 1-13
  • pH stability (short term): 1-14
  • Hydrodynamic resistance: 4 psi backpressure at 1.5 l/min for water at 25 oC

pH Flow cell

  • Total volume: 500μl
  • Flow rates: 0.1 to 1500ml/min, no split
  • Maximum operating pressure: 5 psi with the actual pH electrode, 15 psi with the dummy electrode
  • Wetted materials: stainless 316, Viton, PEEK
  • Solvents compatibility is limited by the pH electrode
  • Hydrodynamic resistance: 3 psi backpressure at 1.5 l/min for water at 25 oC

Monitoring of UV absorption at 250 and 280 nm

  • Range: 0.001 – 100* AU (Absorption Units)
  • Resolution: 0.0001 AU
  • Baseline noise: < 0.0002 AU
  • Drift: 3.0 x 10-4 AU/hr at constant temperature
  • Auto-zero: the detector may be zeroed at any time

Monitoring pH

  • Electrical connector: BNC
  • Input current: <1pA
  • Input voltage range -1500 to +1500 mV, will accept any type of potentiometric pH or Ion Selective electrode
  • Electrode resistance measurement range: 1-5000 MOhm

Monitoring of Conductivity in S/m

  • Two electrodes platinum cell
  • 0.001-15 S/m range

Monitoring of Proton mobility

  • This technology was developed by REACH Devices.
  • Works in any solvent from water to hexanes
    • In aqueous solvents, this channel measures conductivity.
    • In non-aqueous solvents, this channel measures proton mobility: very sensitive for weakly ionic compounds (e.g., allows the user to monitor sugars in dichloromethane).

Data storage

Data is saved automatically to the internal non-volatile memory. The memory can hold a maximum of 262128 data points.

Acquisition rates (non-stop run times)

  • 5Hz (14 hours 33 min)
  • 2Hz (36 hours 24 min)
  • 1Hz (72 h)
  • 0.5Hz (145 h)
  • 0.2Hz (15 days)
  • 0.1Hz (30 days)

Data Transfer

By user's command to USB drive. No data acquisition occurs during the data transfer.

Communication Interface

  • In the field firmware upgrade
  • Up to two relays which can be set by the user to react on change in UV absorbtion, pH or conductivity. Allows for direct or reverse control with hysteresis.(250VAC 4A max)
  • One digital fraction count input from a fraction collector; detector can react to intermittent mechanical contact closure and TTL level signal.
  • Two digital outputs which can be programmed by REACH Devices as requested by the user
  • 4 independent analog outputs in 0-1V range (see manual for transfer functions), corresponding to (a) UV 250 nm, (b) UV 280 nm, (c) refractive index and (d) conductivity channels. The output (a) can be reassigned by a user to provide data from any other channel (b, c, or d) and used to provide the respective analog signal for contemporary fraction collectors.
  • All input and output channels are galvanically isolated from mains ground and detector's internal circuitry ground for safety and noise resistance.

Display

  • Dimensions: 5.7” diagonally
  • Resolution: QVGA 320x240
  • Back-light: white LED
  • Protection: hermetically covered with 2 mm glass – the assembly is solvent resistant.

Case material

Anodized titanium, covered with 1 micrometer Teflon – complete resistance to any solvent spills.

Dimensions

9” x 5.25” x 5” (190 x 133 x 127 mm)

Weight

4 lb (<1.8 kg)

Power requirements

  • Embedded internal (USA, Asia, and most other regions) or included external (Canada, Europe) power supply
  • Frequency: 50-60Hz
  • Voltage: 90-240V
  • Current rating: 0.1A for 90-120V, or 0.06A for 220-240V; 10W max power draw

*The data is simultaneously collected for optical path-lengths 0.001 to 1 mm, and presented to user “as if” the acquisitions occurred using a 1 mm optical path-length. Unlike other UV monitors, the chromatogram data collected by RD4 detector does not plateau when concentrations of UV-absorbing compounds are high, allowing the user to see true analyte peaks as they leave the column.